Technical Data
Ambient conditions
During operation
Temperature
15 to 35
°
C
Atmospheric pressure
700 to 1100 hPa
Relative humidity
0 to 95 %
Magnetic induction (e.g. with scanners)
<70 Millitesla
During storage (not filled, wick dry)
Temperature
–20 to 70
°
C
Atmospheric pressure
500 to 1200 hPa
Relative humidity
0 to 95 %
Magnetic induction (e.g. with scanners)
<70 Millitesla
Set values
Calibrated markings on handwheel
Vapor standard design 4 (5) vol.%
0.2 / 0.4 / 0.6 / 0.8 / 1 / 1.5 / 2 / 2.5 / 3 / 3.5 / 4 / (5) vol.%
Vapor special design 7 (8) vol.%
0.3 / 0.5 / 0.7 / 1 / 1.5 / 2 / 2.5 / 3 / 3.5 / 4 / 5 / 6 / 7 / (8) vol.%
Accuracy of concentration delivered for single
parameter variation
Vapor standard design 4 (5) vol.%
±
0.15 vol.% at 1013 hPa or
±
15 % of set value,
whichever is the greater
Vapor special design 7 (8) vol.%
±
0.2 vol.% at 1013 hPa or
±
20 % of set value,
whichever is the greater
under the following conditions:
– variation of Air flow in flow range given at 22
°
C and 1013 hPa
(measured according to ISO 5358) or
– variation of temperature in range given at an Air flow of
4 L/min and 1013 hPa or
– variation of atmospheric pressure in range given at an Air flow of
4 L/min and 22
°
C
Fresh gas flow range
250 mL/min to 15 L/min
250 mL/min to 10 L/min for concentrations >5 vol.%
Quality of gases required
Clean, medically pure mixtures of O
2
and Air, O
2
and N
2
O
O
2
and Air: dew point
≤
5
°
C at 5 bar
N
2
O: water content
≤
2 mg/L at 5 bar
Flow resistance (without connector)
Vapor setting »0«
<15 mbar at 10 L/min Air, 15 to 35
°
C
Vapor switched on
<60 mbar at 10 L/min Air, 22
°
C
<110 mbar at 10 L/min Air or O
2
, 15 to 35
°
C
27
Technical Data