50
| PT5300
DIAGNOSE/Keyboard
2NDMNU:../DIAGNOSE/Keyboard
Press button:
xxxxx
>>>>>
Description:
The keyboard test is testing the response of the buttons on the front of the instrument. To perform the test press the
button indicated.
The xxxxx will be replaced by a button name in the following sequence: LEFT, DOWN, RIGHT, EXECUTE, PRESET,
OUTPUT, C.BAR, WINDOW, SPECIAL, LINEARITY and PATTERN.
The requested button has to be pressed within a given time. The arrows to the left in the display indicate the time to
press the button, i.e. the button should be pressed before five arrows are visible.
The display will change as soon as the correct button has been pressed or a time-out occurs.
To restart the keyboard test press the E button. If a time-out occurs the key will be marked as FAIL. To return to the
previous menu level, press the
button.
Note:
The is NOT tested in this sequence. It is tested when it is used to exit the menu.
DIAGNOSE/Memory
2NDMNU:../DIAGNOSE/Memory
Testing memory: ROM
>>>>>
Description:
There are two types of memory to test:
•
ROM
•
RAM
When the test is running, this will be indicated by a number of arrow on the right side of the display. When a total of
five arrows are displayed, the test is finished. If no errors have been found, a message OK will appear on the display
to indicate that the test has been completed; otherwise the message FAIL will appear. When the ROM has been
tested the check sum is also displayed.
To continue testing on the next main section, press the E button. To cancel further testing, press the
button.
When both memory types have been tested, a message stating the result of the complete test will appear.
The RAM check is done by writing and reading the sequence Ox55 and OxAA to/from each memory position in the
RAM. Since this operation involves moving the content out of the RAM, there is a potential risk of losing that data if
the apparatus is switched off during the test. A built-in safety procedure will detect such an error during power-up and
will report a general failure in the RAM. If this happens, a factory reset will be performed.
To perform the main test once more, press the E button. To return to the previous menu level, press the
button.
DIAGNOSE/Configuration
The Diagnose Configuration displays the units detected during power up. The identification may include type number
(PT5300), Serial Number (KUxxxxxx), and when available the software version.
Summary of Contents for PT5300
Page 6: ......
Page 59: ...PT5300 53 4 9 Menu Tree Figure 7 PT5300 Menu tree ...