PPCK+™ Operation and Maintenance Manual
Page 85
©2001
DH Instruments, Inc
.
HS%
Purpose
Read or set the hold limit for automated pressure control in percent of range
full scale.
Command
Query
“HS%
n
”
“HS%?”
Classic “HS%=
n
“HS%”
Arguments
N:
The hold limit in % FS of the current active range.
Remarks
The hold limit can be read and set as a pressure or as a percent of the range.
Example
(enhanced)
Cmd sent:
Query reply:
“HS .01”
“0.0100 %”
Example
(classic)
Sent:
Reply:
“HS=.01”
“0.0100 %”
Errors ERR#
6:
The
‘n’
argument was invalid.
See Also
3.1.2.3, 3.1.2.4, 3.1.2.5, 3.2.3, 3.2.6, “HS”
IF
Purpose
Increase the pressure at the fast rate.
Command “IF
n
”
Classic “IF=
n
”
Arguments
N
‘0’
Stops the pressure increase.
‘1’
Starts the pressure increase.
Remarks
PPCK+ will set and maintain the fast increase pressure rate. Care must be used, as
the pressure will not stop increasing until “IF=0” or “ABORT” is sent, or the upper limit
is passed.
Example
(enhanced)
Cmd sent:
Reply:
“IF 1”
“1” (no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“IF=1”
“IF=1”
Errors ERR#
6:
The
n
argument is a ‘0’ or a ‘1’.
See Also
3.1.2.2, “ABORT”, “DF”, “DS”, “DF”, “UL”, “RATES”, “RS”
IP
Purpose
Increase the pressure a given amount.
Command “IP
n
”
Classic “IP=
n
”
Arguments
N
The increase in pressure desired (current pressure units).
Remarks
PPCK+ will increase the pressure by the specified amount and stop.
Example
(enhanced)
Cmd sent:
Reply:
“IP 2”
“2.000 kPa”(no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“IP=2”
“2.000 kPa”
Errors ERR#
6:
The
n
argument is not within given limits.
See Also
“DP”
Summary of Contents for PPCK PLUS
Page 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Page 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Page 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Page 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...