5.
M
AINTENANCE
,
A
DJUSTMENTS
A
ND
C
ALIBRATION
5.1
OVERVIEW
PPCK+ was designed for maintenance free operation. No maintenance is required other than:
•
Rezeroing of reference transducer (RPT) as needed (see Section 3.4.1).
•
Regular RPT recalibration (see Section 5.2).
•
Adjustment of the on-board barometer as needed (see Section 5.3).
•
Purging of the DRAIN port (see Section 5.4).
This section provides information on maintenance, adjustment and calibration procedures, certain
repair functions and recommended overhaul procedures.
PPCK+ is a sophisticated pressure setting and measuring instrument with advanced on-board
features and functions. Before assuming that unexpected behavior is caused by system defect or
breakdown, use this manual and other training facilities to become thoroughly familiar with
PPCK+ operation. For rapid assistance in specific situations (see Section 6).
PPCK+ is covered by a limited one year warranty. Unauthorized service or repair during the warranty
period is undertaken at the owner's risk and may cause damage that is not covered under warranty
and/or may void the warranty.
Summary of Contents for PPCK PLUS
Page 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Page 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Page 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Page 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...