PPCK+™ Operation and Maintenance Manual
Page 135
©2001
DH Instruments, Inc
.
7.
A
PPENDIX
7.1
DRIVERS
The PPCK+ drivers option provides eight open collector drivers for operating external valves,
solenoids, indicators, etc. When operating from the setup-driver screen (see Section 3.3.4), pressing
enter will allow the
operating
mode of the drivers to be set. The two modes of operation are
Momentary and Toggle. A momentary driver will change state while the corresponding driver number
on the keyboard is being pressed. In
toggle
mode, the driver state will toggle each time the
corresponding key is pressed and released.
Each output can sink 500 mA at 12 V. However, the total output of all the activated drivers cannot
exceed 1 Amp. Therefore, if multiple drivers are being activated please refer to the following chart as
a guide.
Table 19.
Driver Output
# OF ACTIVE
DRIVERS
MAX CURRENT
PER OUTPUT
1 500
mA
2 400
mA
3 275
mA
4 200
mA
5 160
mA
6 135
mA
7 120
mA
8 100
mA
The male connector (P/N 401382) for the DRIVERS port is an optional PPCK+ accessory.
Summary of Contents for PPCK PLUS
Page 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Page 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Page 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Page 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...