PPCK+™ Operation and Maintenance Manual
©2001
DH Instruments, Inc
.
Page 114
∗
STB
Purpose
Read the Status Byte Register.
Command
“
∗
STB?”
Remarks
The Status Byte Register reflects the general status of the PPCK+. The ‘MSS’ bit
state is represented by bit 6.
Example
(enhanced)
Sent:
Reply:
“
∗
STB?”
“80” (The MSS and MAV bits are set)
RSE
Purpose
Read or set the Ready Status Enable Register.
Command
Query
“RSE
n
”
“RSE?”
Default “RSE
0”
Arguments
n:
’0 to 255’ This is the decimal representation of the bit(s) to enable.
To enable the RDY bit, the argument would be 1.
Remarks
The Ready Status Enable Register determines which bits in the Ready Status Register
are enabled and included in the Status Byte Register (RSR bit), and can assert the
SRQ line. The reply is in decimal numeric form.
Example
(enhanced)
Sent:
Query reply:
“
∗
RSE=1” (enables the RDY bit)
“1” (no reply if IEEE-488)
Errors ERR#
6:
n
is not valid.
RSR
Purpose
Read the Ready Status Register.
Command “RSR?”
Remarks
The Ready Status Register contents are cleared after reading. The reply is in
decimal numeric form.
Example
(enhanced)
Sent:
Reply:
“RSR?”
“6” (The MEAS and NRDY)
Summary of Contents for PPCK PLUS
Page 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Page 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Page 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Page 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...