ATM
Purpose
Reads the next measured internal ambient pressure.
Query
Classic
“ATM?”
“ATM”
Remarks
The atmospheric pressure as measured by the PPCK+ on-board barometer will be
returned in the current pressure units (always absolute). This measurement is
followed by the units text.
Example
(enhanced)
Query sent:
Query reply:
“ATM?
“97.123 kPaa”
Example
(classic)
Sent:
Reply:
“ATM
“97.123 kPaa”
See Also
3.4.4.7
AUTOZERO
Purpose
Read or set the status of the automatic zero function.
Command
Query
“AUTOZERO
n
”
“AUTOZERO?”
Classic “AUTOZERO=
n
”
“AUTOZERO”
Default
“AUTOZERO 1”
Arguments
n:
’0’ Autozero OFF
’1’ Autozero ON
Remarks
The PPCK+ autozero function can be turned ON and OFF.
Example
(enhanced)
Cmd sent:
Query reply:
“AUTOZERO 1”
“0” If autozero OFF
“1” autozero ON
Example
(classic)
Sent:
Query reply:
“AUTOZERO=1”
“AUTOZERO=1” If autozero ON
Errors
ERR# 6:
The argument was other than a ‘0’ or a ‘1’.
See Also
3.4.1
Summary of Contents for PPCK PLUS
Page 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Page 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Page 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Page 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...