PPCK+™ Operation and Maintenance Manual
©2001
DH Instruments, Inc
.
Page 26
Ready/Not Ready In Static Control Mode
When static pressure control mode is active a
Ready
condition occurs whenever:
•
The current measured pressure is inside the hold limit.
•
The rate of change of pressure is less than the current stability limit.
Figure 8.
Ready
/
Not Ready
in Static Control Mode
Ready/Not Ready In Dynamic Control Mode
When
dynamic pressure control
mode is active, a
Ready
condition
occurs whenever:
•
The current measured pressure is inside the hold limit.
Figure 9.
Ready
/
Not Ready
in Dynamic Control Mode
Not Ready:
Pressure out of
hold limit
Time
Target Pressure
Ready:
Pressure inside of hold limit
P
r
e
s
s
u
r
e