PPCK+™ Operation and Maintenance Manual
Page 25
©2001
DH Instruments, Inc
.
3.1.2.5
PRESSURE READY /NOT READY
PPCK+ provides a
Ready/Not Ready
indication. This indication is intended to
provide the user with a clear and objective criterion for determining when a valid
pressure measurement can be made. On the main run screen, the far left character
of the top display lines indicates
Ready
/
Not Ready
. In remote command responses,
Ready
/
Not Ready
is indicated by “R” or “NR” in the response string.
Ready
/
Not Ready
front panel display indications are:
•
<*>:
Pressure
Ready
.
•
<
↓
>:
Pressure
Not Ready
and decreasing.
•
<
↑
>:
Pressure
Not Ready
and increasing.
When pressure control is NOT active:
A
Ready
condition occurs any time the
pressure rate of change is inside the stability limit. The stability limit is defined in
terms of rate of change of pressure in current pressure units per second (see
Section 3.2.6).
When pressure control is active:
The criteria for determining the
Ready
/
Not
Ready
condition depend on whether the current
control
mode
is
static
or
dynamic
.
Pressure
Ready
/
Not Ready
parameters are set by default when a
control
mode
is selected. The parameters can be customized if desired (see Sections 3.2.3
and 3.2.6).
The Ready condition is handled specially when responding to an automated pressure
command of zero (see Section 3.2.10.1).
Summary of Contents for PPCK PLUS
Page 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Page 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Page 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Page 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...