PPCK+™ Operation and Maintenance Manual
©2001
DH Instruments, Inc
.
Page 96
RES
Purpose
To read or set the pressure display resolution for the current RPT and range.
Command
Query
“RES
n
”
“RES”
Classic “RES=
n
”
“RES”
Default “RES
0.001”
Arguments
n:
The pressure display resolution in % FS of the current RPT range
(0.0001 to 1 % FS).
Remarks
The pressure display resolution is defined as % FS of the selected RPT range.
The setting is separate for each range, and will change as the range is changed.
Example
(enhanced)
Cmd sent:
Query reply:
“RES .01”
“0.01 %FS”
Example
(classic)
Sent:
Reply:
“RES=.01”
“0.01 %FS”
Errors
ERR# 6:
The argument is invalid.
See Also
3.2.9
RESET
Purpose
Reset the user’s settings to factory defaults.
Command “RESET”
Classic “RESET”
Remarks
The PPCK+ has user settings (units, resolution, control modes, etc.) that can be reset
to factory defaults. Communications settings will not be affected. The remote
“RESET” program message corresponds to the local “Reset-sets.” function (see
Section 3.4.4.5, Reset-Sets). The reset cycle takes up to 10 seconds to complete.
Remote communications should not take place during this period.
Example
(enhanced)
Cmd sent:
Reply:
“RESET”
“RESET” (no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“RESET”
“RESET”
See Also
3.4.4.5
Summary of Contents for PPCK PLUS
Page 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Page 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Page 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Page 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Page 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...