28
4.6 Firing and pressing programs
The following is an illustrative example of how to execute a firing program or pressing programs
(Multimat® Cube press only).
4.6.1 Overview of firing and pressing parameters
Starting temperature
Specifies the temperature at which the firing/pressing program starts.
Pre-drying time
Specifies the time during which the starting temperature will be held with the firing
chamber open.
Closing time
Specifies the time during which the firing chamber will close gradually; the starting
temperature will he held.
Pre-heating time
Specifies the time during which the starting temperature will be held with the firing
chamber closed.
Vakuum
Specifies the vacuum level in the firing chamber.
Firing stages
Specifies the number (max. 3) of stages (holding levels) used.
Heating rate
Specifies the speed of the temperature rise.
End temperature
End temperature reached during the firing/pressing program.
Vacuum time
Specifies the time during which the vacuum will be maintained after the end tem-
perature is reached.
Holding time
Specifies the time during which the final temperature will be held.
Tempering temperature
Specifies the tempering temperature to be reached.
Tempering time
Specifies the time during which the tempering temperature will be held.
Cooling
Specifies the time during which the firing chamber will slowly open
Pressdruck
Specifies the pressure exerted on the pressing plunger.
Pressing time
Specifies the time during which the pressing ie performed.