33
Error message
Possible causes
Possible remedies
Vacuum system error.
Program must be aborted.
Vacuum system is not ventilated.
Vacuum cannot be relieved.
Contact Dentsply Sirona.
Vacuum timeout.
Program must be aborted.
Target vacuum level is
not reached.
Chamber seal or
vacuum system leak.
Perform vacuum test.
Check the chamber seal and
vacuum pump,
Section 3.3.
If the error recurs, contact
Dentsply Sirona.
A system error has occurred.
Try to restart the program.
Error code: 0x407
Error in system communication.
Restart the program.
If the error recurs, contact
Dentsply Sirona.
Lift error.
Limit switch defective.
Contact Dentsply Sirona.
Supply voltage failed while firing
was in progress. Firing cannot be
resumed.
Supply voltage failed briefly while
the program was executed.
The vacuum and/or temperature
were no longer sufficient to conti-
nue the program.
Check the firing/pressing result.
Supply voltage failed while firing
was in progress. However, the
program was completed.
Supply voltage failed briefly while
the program was executed.
Check the firing/pressing result.
Battery error #1
The battery of the control unit
is empty.
Contact Dentsply Sirona.
Battery error #2
The battery of the control board
is empty.
Contact Dentsply Sirona.
Export failed
Error exporting data from unit to
USB stick.
Insufficient memory on USB stick.
USB stick defective.
Check the USB stick.
Self-test failed:
Lift problem
Final position not reached: Lift
blocked or limit switch defective.
Cycle the unit ON and OFF,
Section 4.1.
If the error recurs, contact
Dentsply Sirona.
Self-test failed:
Vacuum problem
No vacuum present.
Vacuum pump not connected
or defective.
Connect vacuum pump,
Section 3.3.
Check connections.
If the error recurs, contact
Dentsply Sirona.
Self-test failed:
Pressing system problem
Pressing plunger could not be
extended.
Connect to compressed air,
Section 3.3.
Check connections.
If the error recurs, contact
Dentsply Sirona.
Self-test failed:
Temperature problem
Temperature does not increase.
Heater or thermocouple defective.
Cycle the unit ON and OFF,
Section 4.1.
If the error recurs, contact
Dentsply Sirona.
Summary of Contents for Multimat Cube
Page 1: ...Instruction Manual Multimat Cube Multimat Cube press...
Page 39: ...39...