Calibrating:
Open the firing chamber.
Place the carrier with the silver wire in the middle
of the firing base.
Touch <Start>.
After the program has run, a prompt will appear
asking whether the calibration value should be
accepted.
Before proceeding, the result of the calibration
carried out should first be evaluated as follows:
One or more small beads have
formed on the silver wire.
= Temperature correct
= Temperature too high!
= Temperature too low!
The silver wire has almost or
completely melted.
No melting effect is visible
on the silver wire.
Bewertung:
Weiteres Vorgehen:
Temperature correct
Touch <Yes>. The calibration value is automatically stored under
<Settings>
<Furnace>. This concludes the calibration of the unit.
Temperature too high
Touch <No>. The calibration value will not be stored.
Correct the final temperature value downwards in increments
of approx. 3°C.
Touch <Start> to restart the program.
Temperature too low
Touch <No>. The calibration value will not be stored.
Correct the final temperature value downwards in increments
of approx. 3°C.
Touch <Start> to restart the program.