Datex-Ohmeda S/5 monitors
50
Document No. 8005571-1
TROUBLE CAUSE TREATMENT
Out of range
Measured pressure is beyond the
internal measurement range of
the module.
The waveform hits the top and the
numeric display not shown. Check
transducer and its level. Zero the
channel.
4.1.7 Impedance respiration
TROUBLE CAUSE
TREATMENT
No resp trace
Waveform not selected on the
screen
Press the Monitor Setup key and
make adjustments.
Module not plugged in correctly
Re-plug the module.
Unacceptable resp waveform
Poor electrode or poor electrode
skin contact
Electrodes from different
manufacturers are used. Too
much/little gel is used.
Poor electrode condition
Electrodes are dried out.
Improper site of electrodes
Check that electrodes are not placed
over bones, active muscles, or layers
of fat.
Improper skin preparation
Remove body hair. Clean
attachment site carefully with
alcohol.
Message: ‘SMALL RESP CURVE’
Respiration signal is very small
With 3-lead cable in ESTPR/NESTPR
try another lead connection I, II, III or
try 5-lead cable.
Message: ‘APNEA ALARM’, and
respiration waveform normal
Respiration source is CO
2
Check respiration source and
change it to correct one.
Summary of Contents for S/5 PRESTN
Page 2: ......
Page 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Page 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Page 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Page 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Page 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...