S/5 M-PRESTN Modules
5
Document No. 8005571-1
±
0.2 °C (10 to 24.9 °C)
Display resolution
0.1 °C (0.1 °F)
Temperature test
automatic (every 10 min)
Probe type
compatible with YSI 400 series
Single use sensors
±
0.2 °C (25 to 45.0 °C)
±
0.3 °C (10 to 24.9 °C)
1.2.5 Invasive blood pressure
Measurement range
-40 to 320 mmHg
Measurement accuracy
±
2 mmHg or
±
5 %
Zero adjustment range
±
150 mmHg
Calibration range
±
20 %
Scales
upper limit is adjustable between 10 and 300 mmHg in steps of
10. Lower limit is 10 % of selected upper limit below zero.
Sweep speed
12.5, 25, 50 mm/s
DIGITAL DISPLAY
Range
-40 to 320 mmHg
Resolution
±
1 mmHg
WAVEFORM DISPLAY
Range
-30 to 300 mmHg
PULSE RATE FROM ARTERIAL PRESSURE
Measurement range
30 to 250 bpm
Resolution 1
bpm
Accuracy
±
5 bpm or
±
5 % whichever is greater
1.2.6 Respiration
NOTE: The respiration measurement is intended for patients over three years old
Measurement range
4 to 120 bpm
Accuracy
±
5 bpm or
±
5 %
Resolution 1
bpm
Averaging time
30 s
Update interval
10 s
RESPIRATION WAVEFORM
Sweep Speeds
6.25 mm/s and 0.625 mm/s