5.1.9 ECG service menu
Power freq
Set power frequency; 50 Hz/60 Hz.
Filter low
Set filter low frequency; 0.05 Hz/0.5 Hz.
Filter high
Set filter high frequency; 30 Hz (40 Hz if power freq is 60 Hz) /100 Hz or 150 Hz @ NE12STPR.
Service Data field
Power freq, and Cable type
show the values chosen or detected,
Filter low and high
defines the
selected filter (Monitor/Diagnostic/ST).
Quick zero
at PRESTN module is ON when the ECG signal is beyond scale, and therefore, is quickly
returned to optimal range using fast signal processing methods. All the
Quick zero
bits are ON at
the same time.
Cable
shows ON when ECG cable is connected.
Electrode
shows ON when each of these electrodes are connected.
Pacer count
is a running number for pacemaker users.
Button
No effect on the module.
Resp Available
indicates that ECG hardware is capable of measuring impedance respiration.
Measurement
shows ON when the respiration measurement is on.
Amp zero
shows
ON when zeroing of the respiration amplifier takes place.
Waveform
VALUE
will be updated in one second interval.
Timeouts
is a cumulative number that indicates how many times the module has not responded to
the monitor’s inquiry.
Bad checksums
is a cumulative number that indicates how many times
communication from the module to monitor broke down.
Bad c-s by mod
is a cumulative number that indicates how many communication errors the module
has detected.
Summary of Contents for S/5 PRESTN
Page 2: ......
Page 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Page 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Page 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Page 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Page 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...