![Datex-Ohmeda S/5 PRESTN Technical Reference Manual Slot Download Page 66](http://html.mh-extra.com/html/datex-ohmeda/s-5-prestn/s-5-prestn_technical-reference-manual-slot_513816066.webp)
Datex-Ohmeda S/5 monitors
60
Document No. 8005571-1
5.1.7 NIBP pneumatics menu
Start Pump/Stop Pump
A manual control for the pump. The selection changes to
Stop Pump
when the pump turns on.
Open Exh1/Close Exh1
No effect on the module.
Open Exh2/Close Exh2
A manual control for the exhaust valve 2. The selection changes to
Close Exh2
when the valve is
opened.
Open Zero valve
No effect on the module.
Set Valve
With
Set Valve
, the opening of the pulse valve is adjusted between 0 and 255 (0 for fully closed
and 255 for fully open). First push the ComWheel, then turn it to adjust the value on screen and
finally push to set the value.
Reset Clock
Reset Clock
will zero the time on the display.
Pneumatics Data field
See NIBP service menu in chapter
for information on general items
Pressure, Zero, Protect
handle, Calibr. prot., +15 V power, AD0
to
AD7
as well as
Timeouts
etc.
Pump, Exh1 Valve
, and
Exh2 Valve
show their states. Note that
Exh1 Valve
has no effect to the
module.
Pulse Valve
shows how much the valve is opened (0 to 255) during Valve Setting.
Interval 20 mmHg -> 185 mmHg Checking
Select the
Start pump
at different combinations of the valves open/closed and push the
ComWheel. The module counts the time it takes for the pressure to go up from 20 mmHg to 185
mmHg and displays it. When all the valves are closed, the pump should be able to pump the
pressure in about 1 to 4 seconds into an adult cuff wrapped around a pipe. The pump does not
stop without selecting the
Stop Pump
by pushing the ComWheel.
Summary of Contents for S/5 PRESTN
Page 2: ......
Page 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Page 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Page 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Page 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Page 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...