![Datex-Ohmeda S/5 PRESTN Technical Reference Manual Slot Download Page 57](http://html.mh-extra.com/html/datex-ohmeda/s-5-prestn/s-5-prestn_technical-reference-manual-slot_513816057.webp)
S/5 M-PRESTN Modules
51
Document No. 8005571-1
4.2 Troubleshooting flowcharts
4.2.1 M-PRESTN module troubleshooting for NIBP parameter
Possible faulty
module with NIBP
parameter.
Remove all
modules from
Monitor Frame.
Insert NXXX module
and turn power on.
Does fault
still appear?
Fault not in
NXXX module.
Module ID
on screen?
Does
another module
work in same
slot?
Error
message on
screen?
See error code explanation
in Service Manual and
fix it.
Do
they work?
Check tubes and
connectors. Find leak and
fix it.
Check keyboard
connector. If OK,
change front panel.
Check and
replace
NIBP board.
OK?
NO
YES
YES
NO
NO
NO
YES
NO
NO
YES
NO
YES
Key check
in Service Menu
OK?
Pump check
in Service Menu
OK?
Leak test
in Service Menu
OK?
YES
NO
Check pump
connector. If OK,
change pump.
YES
START
NO
OK?
Replace
NXXX module.
Check and
replace
NIBP board.
NO
N
E
S
T
P
R
_modul
e_troubl
es_fl
o
w
ch.vsd
Figure 18
M-PRESTN module troubleshooting flowchart for NIBP Parameter
Summary of Contents for S/5 PRESTN
Page 2: ......
Page 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Page 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Page 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Page 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Page 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...