S/5 M-PRESTN Modules
57
Document No. 8005571-1
2. For proper zeroing to take place, remove the hose from the front panel connector. Select
Calibration
and push the ComWheel. Messages ‘Zeroing’ and ‘Zeroed’ will appear in the NIBP
message field. After this a pressure bar will appear beside the menu and the text ‘Calibrating’
will appear to NIBP digit field.
3. Connect an external mercury manometer with pump to module through the both tubes of the
hose. Pump up to about 200 mmHg pressure (range of 150 to 300 mmHg allowed) according
to the manometer. Verify that both pressure values in the prompt field match the manometer
reading. If not, adjust by turning the ComWheel.
4. When the values are equal, push the ComWheel to confirm the calibration. After a few
seconds ‘Calibrated’ message will appear to the NIBP digit field, which means that the
calibration data has now been saved.
NOTE: When calibrating NIBP, always change the displayed pressure value slightly with the
ComWheel, even in cases where the value would be correct. For example change the value one step
higher and then back one step lower. “Calibrated” text should appear in the display. This ensures
that the calibration procedure is correctly registered and stored by the module.
5. Use the module buttons again to enable
Protection
setting and set it ON, and finally disable
Protection
setting.
5.1.4 NIBP safety valve menu
Start Test
Start test
is for starting and
Stop test
is for stopping the Safety Valve test.
Safety Valve Data
See NIBP Service menu in chapter
for information on general items
Pressure, Zero, Protect
handle, Calibr. prot., +15 V power, AD0
to
AD7
as well as
Timeouts
etc.
Max. press
and
2 s after stop
show the measured values at Safety Valve test.
Safety Valve Test Adult/Infant
Wrap an adult cuff around a pipe and connect the cuff to the module. Highlight
Start test
and give
the ComWheel a push. The test ends automatically or when
Stop test
(appears in place of
Start
test
) is pushed.
Summary of Contents for S/5 PRESTN
Page 2: ......
Page 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Page 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Page 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Page 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Page 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...