1.3 Technical specifications
1.3.1 NIBP
Deflation rate, PR dep.
3 to 8 mmHg/s
Inflation time
20 to 185 mmHg, 1 to 5 s
Automatic software control, max. inflation pressure
adult
280
±
10 mmHg
child
200
±
10 mmHg
infant
145
±
5 mmHg
Over pressure limit, stops measurement after 2 seconds
adult
320
mmHg
child
220
mmHg
infant
160
mmHg
The safety circuit limits the maximum cuff pressure to 320 mmHg in adult/child mode or 160
mmHg in infant mode. Independent timing circuit limits pressurizing (>15 mmHg) time to 3 minutes
maximum in adult/child mode, and 90 seconds at (>5mmHg ) in infant mode.
Zeroing to ambient pressure is done automatically.
Inflation pressure is adjusted according to the previous systolic pressure, typically 40 mmHg
above. If the systolic pressure is not found, inflation pressure is increased typically 50 mmHg.
Max. measurement time
adult
120 s
child
120 s
infant
75
s
Pressure transducer accuracy is better than
±
3 mmHg or
±
2 % whichever is greater.
Max. error
±
4 mmHg.
Protection against electrical
shock
Type BF defibrillation proof
1.3.2 ECG
Defibrillation protection
5000 V, 360 J
Recovery time
5 s
Input impedance
>2.5 M
Ω
(10 Hz)
CMRR
≥
95 dB (ST)
System noise
<30
µ
V (p-p, RTI)
Allowable offset
±
800 mVDC
Gain range
0.2 to 5.0 cm/mV
Pacemaker pulse detection
2 to 700 mV, 0.5 to 2 ms pulses
Protection against electrical
shock
Type CF defibrillator proof
1.3.3 Pulse oximetry
Protection against electrical
shock
Type BF defibrillation proof
Summary of Contents for S/5 PRESTN
Page 2: ......
Page 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Page 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Page 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Page 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Page 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...