S/5 M-PRESTN Modules
29
Document No. 8005571-1
3 SERVICE
PROCEDURES
3.1 General service information
Field service of the hemodynamic modules is limited to replacing faulty printed circuit boards or
mechanical parts. Faulty printed circuit boards should be returned to Datex-Ohmeda for repair.
Datex-Ohmeda is always available for service advice. Please provide the unit serial number, full
type designation, and a detailed description of the fault.
Only trained personnel with appropriate equipment should perform the tests and repairs outlined in
this section. Unauthorized service may void warranty of the unit.
3.2 Service check
These instructions include complete procedures for a service check. The service check should be
performed after any service repair; however, the service check procedures can also be used for
determining possible failures.
The procedures should be performed in ascending order.
The instructions include a check form , which should be filled in when performing the procedures.
The mark
"
in the instructions means that the performed procedure should be signed in the
check form.
3.2.1 Recommended tools
Tool Order
No.
Notes
Patient simulator
-
Pressure manometer
-
Temperature test set
884515
Multi-link 3-leadwire set
Multi-Link 5-leadwire set V2-V6
Multi-Link 5-leadwire set C2-C6
Multi-Link 12-lead ECG trunk cable
SpO
2
finger probe
OXY-F4-N or SAS-F4
InvBP transducer
Adult NIBP cuff & hose
Infant NIBP cuff & hose
Screwdriver
3.2.2 Recommended parts
Part Order
No.
Notes
NIBP pump filter
57142
Summary of Contents for S/5 PRESTN
Page 2: ......
Page 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Page 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Page 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Page 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Page 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...