1 SPECIFICATIONS
1.1 General specifications
Module size
75
×
180
×
112 mm
W
×
D
×
H
3.0
×
7.1
×
4.4 in
Module weight
0.7 kg / 1.5 lbs
Power consumption
about 6 W
Operation temperature
10 to 40
°
C / 50 to 104
°
F
1.2 Typical performance
1.2.1 NIBP
NOTE: Non-invasive blood pressure measurement is intended for patients weighing over 5 kg (11
lb.)
Oscillometric measurement principle.
Measurement range
adult
25 to 260 mmHg
child
25 to 195 mmHg
infant
15 to 145 mmHg
Pulse rate range accepted
30 to 250 bpm
Measurement interval
STAT (continuous 5 min), 1, 2.5, 3, 5, 10, 15, 30 and 60 min (1 h),
2 and 4 h
Typical measuring time
adult
23 s
infant
20
s
Initial inflation pressure
adult
170
±
10 mmHg
child
150
±
10 mmHg
infant
120
±
10 mmHg
Venous stasis
adult
40
±
5 mmHg / 2 min
child
40
±
5 mmHg / 2 min
infant
30
±
5 mmHg / 1 min
Cuff widths
please see
User’s Guide
1.2.2 ECG
Lead selection, 12-lead ECG
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Lead selection, other modules
I, II, III, aVR, aVL, aVF, V
Sweep speeds
12.5, 25, 50 mm/sec
DISPLAY FILTER
Diagnostic, 12-lead ECG
0.05 to 150 Hz
Diagnostic, other modules
0.05 to 100 Hz
Monitoring
0.5 to 30 Hz (-3 dB, with 50 Hz reject filter)
0.5 to 40 Hz (-3 dB, with 60 Hz reject filter)
Summary of Contents for S/5 PRESTN
Page 2: ......
Page 6: ...Datex Ohmeda S 5 monitors iv Document No 8005571 1 ...
Page 77: ...S 5 M PRESTN Modules 71 Document No 8005571 1 7 EARLIER REVISIONS No earlier revisions ...
Page 78: ...Datex Ohmeda S 5 monitors 72 Document No 8005571 1 For your notes ...
Page 79: ...S 5 M PRESTN Modules 73 Document No 8005571 1 APPENDIX A ...
Page 80: ...Datex Ohmeda S 5 monitors 74 Document No 8005571 1 ...