2.2.3 ECG board in 12-lead measurement
The 12-lead ECG measurement consists of the elements shown in Figure 6. All functions are
located in the ECG board except the ECG input unit .
PATIENT AND ECG ELECTRODES
ECG CABLE
- ECG LEAD SET
- ECG TRUNK CABLE
ECG INPUT UNIT
- ECG CONNECTOR
- INPUT PROTECTION RESISTORS
INPUT PROTECTION DIODES FOR ECG & RESPIRATION MEASUREMENT
INPUT FILTERING FOR ECG & RESPIRATION MEASUREMENTS
RESPIRATION
MEASUREMENT
AMPLIFIERS
RESPIRATION
MEASUREMENT CURRENT
SUPPLY
ECG PREAMPLIFIERS &
RLD CIRCUIT
LEADS OFF & PACER &
DEFIBRILLATION
DETECTION
BASELINE
RESTORATION
ECG CPU
RS 485 COMMUNICATION
POWER SUPPLY
NV MEMORY
ISOLATION
ISOLATION
MODULE BUS CONNECTOR
ECG BLOCK DIAGRAM
Figure 6
12-lead ECG measurement block diagram