TROUBLESHOOTING
4.1
Troubleshooting charts
See also the
User’s Reference Manual
for more troubleshooting procedures.
4.1.1
NIBP
TROUBLE
CAUSE
TREATMENT
No NIBP value displayed
NIBP not selected on screen.
Check monitor setup.
NIBP menu fading
No M-NE(12)STPR module, module
not properly connected, or NIBP and
NE(12)STPR module connected at
the same time.
Plug in the module.
Artifacts-message
Unsuccessful measurement due to
patient movements or shivering.
Weak pulsation-message
Weak or unstable oscillation pulses
due to:
•
artifacts (accurate diastolic
pressure difficult to measure)
•
marked arrhythmia
•
marked drop in diastolic
pressure
•
diastolic pressure difficult to
measure
•
improper cuff position or
attachment
•
too few pulses detected
•
weak or unusual blood
circulation
•
may give systolic value
Check patient condition and retry.
Check any leaks and retry.
Use proper size of cuff. Check
attachment.
Call service
Error X-message
NIBP hardware error.
X = error number.
See the description of the error message
, the causes and the
solutions listed in the next chapter.