S/5 Hemodynamic modules
Document No. 800 1008-2
29
3.2.2
Recommended parts
Part
Order No.
Notes
NIBP pump filter
57142
All modules
Detach the module box by removing the two screws from the back of the module. Be careful with
loose latch and spring pin for locking.
1.
Check internal parts:
−
screws are tightened properly
−
cables are connected properly
−
all IC’s that are on sockets are attached properly
−
EMC covers are attached properly
−
there are no loose objects inside the module
?
2.
Check external parts:
−
the front cover and the front panel sticker are intact
−
all connectors are intact and are attached properly
−
the module box, the latch and the spring pin are intact
?
3.
Replace the NIBP pump filter in NE12STPR/NE12TPR/NE12STR/NESTPR/NETPR/NESTR
modules, if necessary.
?
•
Reattach the module box and check that the latch is moving properly.
•
Switch the monitor on and wait until the monitoring screen appears. Configure the monitor
screen so that all the needed parameters are shown, for example as follows:
Monitor Setup -
Waveform Fields - Field 1 - ECG1
Field 2 - ECG2
Field 3 - P1
Field 4 - P2
Field 5 - Pleth
Field 6 - Resp
Digit Fields - Lower Field 2 - NIBP
Lower Field 3 - T1+T2