S/5 Hemodynamic modules
Document No. 800 1008-2
33
Temperature measurement
15.
Enter the ESTP : STP service menu:
Parameters -
ESTP : STP
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check also that the STP board memories have passed the
internal memory test, i.e. the ‘RAM’, ‘ROM’ and ‘EEPROM’ show all OK.
?
16.
Check that the ‘Cable’ and ‘Probe’ show OFF for both channels, T1 and T2, when no probes
are connected.
Connect a temperature test plug into the connector T1. Check that the ‘Cable’ and ‘Probe’
for T1 show ON and the corresponding temperature value appears onto the monitor screen.
Perform the same check also for the channel T2.
?
17.
Check the temperature calibrations using temperature test plugs.
If the deviation on a temperature reading on the screen is more than 0.1 °C, calibrate the
temperature channels according to the instructions in the chapter
?
18.
Activate the temperature test by selecting
Temp Test
from the menu and pressing the
ComWheel twice. When the message ‘Performing temp test’ disappears from the digit field,
check that no error messages appear and ‘Temp error’ shows OFF for both channels in the
service menu.
?
19.
Check that the module configuration has been set correctly. The configuration in use is
shown beside the text ‘Configuration’ in the service menu and it can be either STP, ST or TP.
Change the configuration in the
Calibrations
menu, if necessary.
?
Invasive blood pressure measurement
20.
Check the front panel membrane keys that are related to the InvBP or temperature
measurement.
Press each of the keys at least for one second. Check that the pressed key is identified, i.e.
one of the texts for ‘Buttons’ changes from OFF to ON in the service menu.
?
21.
Check that the ‘Cable’ and ‘Probe’ for P1 show OFF. Plug a cable with an invasive blood
pressure transducer into the front panel connector P1 and check that the ‘Cable’ and
Summary of Contents for S/5 M-NE12STPR