S/5 Hemodynamic modules
Document No. 800 1008-2
55
EEPROM
indicates if the values stored in the permanent memory are valid.
The state is either
OK, Fail
or
?
(module not in place or a communication error).
5.1.1
NIBP demo menu
A service menu for demonstrating the oscillometric method of NIBP measurement. The menu
shows the realtime pressure signals that are measured from the NIBP cuff. The measurement result
is shown in the adjoining digit field.
Wave Recording Wave Recording
is for selecting the recording option. If ON is selected, the pressure signals are
recorded in realtime onto the M-REC paper.
Remove menu
Remove menu
widens the displayed waveform area.
Previous Menu
The menu can be closed by selecting the
Previous Menu
or just by pressing the ComWheel if the
Remove menu
was selected.