S/5 Hemodynamic modules
Document No. 800 1008-2
5
1.3
Technical specifications
1.3.1
NIBP
Deflation rate, PR dep.
5...13 mmHg/s
Inflation time
20...185 mmHg, 1...5 s
Automatic software control, max. inflation pressure
adult
280
±
10 mmHg
child
200
±
10 mmHg
infant
150
±
10 mmHg
Over pressure limit, stops measurement after 2 seconds
adult
320 mmHg
child
220 mmHg
infant
165 mmHg
Safety valve limits the maximum cuff pressure to 320 mmHg in adult/child mode or 165 mmHg in
infant mode. Independent timing circuit limits pressurizing (>5 mmHg) time to 2 minutes 10
seconds maximum in adult/child mode, and 1 minute 5 seconds in infant mode.
Zeroing to ambient pressure is done automatically.
Inflation pressure is adjusted according to the previous systolic pressure, typically 40 mmHg
above. If the systolic pressure is not found, inflation pressure is increased typically 50 mmHg.
Max. measurement time
adult
2 min
child
2 min
infant
1 min
Pressure transducer accuracy is better than
±
3 mmHg or
±
2 % whichever is greater.
Max. error
±
4 mmHg.
Protection against electrical
shock
Type BF defibrillation proof
1.3.2
ECG
Defibrillation protection
5000 V, 360 J
Recovery time
2 s
Input impedance
>2.5 M
Ω
(10 Hz)
CMRR
>100 dB (ST)
System noise
<40
µ
V (p-p, RTI)
Allowable offset
±
300 mVDC
Gain range
0.2...5.0 cm/mV
Pacemaker pulse detection
2...500 mV, 0.5...2 ms pulses
Protection against electrical
shock
Type CF defibrillator proof