Datex-Ohmeda S/5 monitors
Document No. 800 1008-2
52
4.2.2
M-ESTPR, M-ESTR, and M-ETPR module troubleshooting
Possibly faulty
ESTPR module
Check module configuration in
Service menu (ESTPR/STP/Calibrations)
Is it
correct?
Select the correct
configuration
Remove all
modules from
the frame
Insert ESTPR module
and turn power on
YES
NO
Does fault
still appear?
Fault not in
ESTPR module
STP
module ID
on screen?
Does
another
module work
in same
slot?
ECG
module ID
on screen?
Check front panel
key functions in
Service menu
Do
they work?
Check front panel unit,
ECG board, and STP board
to find culprit. Replace if
necessary.
Remove
ECG board
STP
module ID
on screen?
Replace
STP board
Replace
ECG board
NO
YES
NO
NO
NO
YES
NO
YES
NO
YES
Has SpO2
measurement
failed?
Replace
STP board
Replace
STP board
YES
NO
YES
Has Temp
measurement
failed?
Replace
ECG board
YES
NO
Has ECG
measurement
failed?
Replace
STP board
YES
Has STP
board failed?
Replace
STP board
NO
NO
YES
Has invBP
measurement
failed?
YES
YES
Enter the
Service menu
Figure 16
M-ESTPR Module Troubleshooting Flowchart