![Datex-Ohmeda S/5 M-NE12STPR Technical Reference Manual Slot Download Page 31](http://html1.mh-extra.com/html/datex-ohmeda/s-5-m-ne12stpr/s-5-m-ne12stpr_technical-reference-manual-slot_3808974031.webp)
S/5 Hemodynamic modules
Document No. 800 1008-2
25
2.3.2
Front panel connectors
Table 4
Front panel connectors
12-lead ECG connector
6
8
9
2
4
5
7
3
1
10
11
12
Pin No
Signal
Pin No
Signal
1
Right arm electrode (R)
7
Chest electrode (C3)
2
Left arm electrode (L)
8
Chest electrode (C4)
3
N
9
Chest electrode (C5)
4
Left leg electrode (F)
10
Chest electrode (C6)
5
Chest electrode (C1)
11
Cable type
6
Chest electrode (C2)
12
Ground
ECG connector (ECG)
SpO
2
connector (SpO
2
)
Invasive blood pressure
connectors (P1, P2)
6
8
9
0
1
2
4
5
7
3
6
8
9
0
1
2
4
5
7
3
6
8
9
0
1
2
4
5
7
3
Pin No
Signal
Pin No
Signal
Pin No
Signal
1
Right arm electrode (R)
1
Feedback resistor
1
Pr
2
Left arm electrode (L)
2
Ground
2
Pressure -
3
Right leg electrode (RL)
3
Not Connected
3
Polarisation - (ground)
4
Left leg electrode (F)
4
Cable
probe identification ground
4
Polari
5
Chest electrode (C)
5
Probe identification
5
Not connected
6
Cable shield
6
LED drive ground
6
Not connected
7
Not connected
7
LED drive current
7
Not connected
8
3/5 lead identification
8
Input signal current
8
Not connected
9
Lead connection check
9
Ground
9
Ground
10
Ground
10
Ground
10
Cable detection
Summary of Contents for S/5 M-NE12STPR
Page 2: ......
Page 6: ...Datex Ohmeda S 5 monitors Document No 800 1008 2 iv This page intentionally left blank...
Page 83: ...APPENDIX A Service check form S 5 Hemodynamic modules Document No 800 1008 2 77 APPENDIX A...