Datex-Ohmeda S/5 monitors
Document No. 800 1008-2
4
Adjustable pulse beep volume.
PLETH WAVEFORM
Scales
2, 5, 10, 20, 50 mod%, Auto
Start up scale is 20 mod% if AUTO is not selected to be the default setting.
1.2.4
Temperature
Measurement range
10...45 °C (50...113 °F)
(In rev. ESTP 03/ EST 02/ETP 02 or earlier: 15...45 °C (59...113 °F))
Measurement accuracy
±
0.1 °C (25...45.0 °C)
±
0.2 °C (10...24.9 °C)
Display resolution
0.1 °C (0.1 °F)
Temperature test
automatic (every 10 min)
Probe type
compatible with YSI 400 series
1.2.5
Invasive blood pressure
Measurement range
-40...320 mmHg
Measurement accuracy
±
2 mmHg or
±
5 %
Zero adjustment range
±
150 mmHg
Calibration range
±
20 %
Scales
upper limit is adjustable between 10 and 300 mmHg in steps of
10. Lower limit is 10 % of selected upper limit below zero.
Sweep speed
12.5, 25, 50 mm/s
DIGITAL DISPLAY
Range
-40...320 mmHg
Resolution
±
1 mmHg
WAVEFORM DISPLAY
Range
-30...300 mmHg
HEART RATE FROM ARTERIAL PRESSURE
Measurement range
30...250 bpm
Resolution
1 bpm
Accuracy
±
5 bpm or
±
5 % whichever is greater
1.2.6
Respiration
Measurement range
4...120 bpm
Accuracy
±
5 bpm or
±
5 %
Resolution
1 bpm
Averaging time
30 s
Update interval
10 s
RESPIRATION WAVEFORM
Sweep Speeds
6.25 mm/s and 0.625 mm/s