S/5 Hemodynamic modules
Document No. 800 1008-2
45
TROUBLE
CAUSE
TREATMENT
Cuff occlusion-message
1. Cuff and/or hose occluded.
Reason:
1.
−
cuff tube kinked
−
straighten tube
−
tube inside module kinked
−
straighten tube
−
occlusion inside/outside
module
−
remove occlusion
2. Cuff, hose, and tubes OK.
Reason:
2.
−
fault in pressure transducer
−
replace the NIBP board
−
fault in A/D converter
−
replace the NIBP board
−
faulty calibration
−
check calibration
−
missing voltages
−
recalibrate
Calibration switch on -
message
EEPROM protection switch at the
bottom of the module is turned to
right.
Enables setting the protection OFF in the
Calibration menu. Turn switch to left if you
are not going to calibrate.
Calibration not protected -
message.
Calibration protection is set to
OFF.
Set the protection ON in the NIBP
Calibration menu.