S/5 Hemodynamic modules
Document No. 800 1008-2
51
4.2
Troubleshooting flowcharts
4.2.1
M-NE12STPR and M-NESTPR module troubleshooting
Possible fault
in NIBP module.
Remove all
modules from
Monitor Frame.
Insert NIBP module
and turn power on.
Does fault
still appear?
Fault not in
NIBP module.
Module ID
on screen?
Does
another
module work
in same
slot?
Error
message on
screen?
See error code explanation
in Service Manual and
fix it.
Do
they work?
Check tubes and
connectors. Find leak and
fix it.
Check keyboard
connector. If OK,
change front panel.
Check and
replace
NIBP board.
OK?
NO
YES
YES
NO
NO
NO
YES
NO
NO
YES
NO
YES
Key check
in Service Menu
OK?
Pump check
in Service Menu
OK?
Leak test
in Service Menu
OK?
YES
NO
Check pump
connector. If OK,
change pump.
YES
START
NO
OK?
Replace
NIBP module.
Check and
replace
NIBP board.
NO
Figure 15
M-NE12STPR and M-NESTPR module troubleshooting flowchart