30
4.
Plug in the module. Check that it goes in smoothly and locks up properly
?
5.
Check that the module is recognized, i.e. all needed parameter information, except invasive
blood pressure, starts to show on the screen.
?
Preset ECG, Respiration, InvBP and SpO
2
measurement settings:
ECG -
ECG Setup - Hr Source - Auto
Pacemaker - Show
Others
- Resp Setup - Size - 1.0
Resp Rate Source - Auto
Measurement - On
Detection Limit - Auto
Invasive Pressures
- P1 ‘Art’ Setup - Label - Art
P2 ‘Cvp’ Setup - Label - Cvp
Pulse Oximetry
- Pleth Scale - Auto
ECG measurement
6.
Enter the service menu:
Monitor Setup
- Install/Service
(password 16-4-34)
-
Service
(password 26-23-8)
- Parameters
Take down the information regarding module software by selecting
Scroll Vers
and turning
the ComWheel.
?
7.
Enter the ESTP : ECG service menu:
Check that the ‘Timeouts’, ‘Bad checksums’ and ‘Bad c-s by mod’ values are not increasing
faster than by 50 per second. Check also that the ECG/RESP board memories have passed
the internal memory test, i.e. the ‘RAM’, ‘ROM’ and ‘EEPROM’ state all OK.
?
8.
Check the front panel membrane key
ECG Lead
(not available in NE12STPR/NESTPR type
modules).
Press the key at least for two seconds. Check that the selected ECG lead is changing on the
screen and the state for ‘Button’ in the service menu.
?