4.1.2
NIBP error code explanation
Code Explanation
Treatment
0
RAM failure; memory failure
Change NIBP board.
1
ROM checksum error; memory failure
Change NIBP board.
2
+15 V failure
Check short circuits. Change NIBP
board.
3
-15 V failure
Check short circuits. Change NIBP
board.
4
EEPROM protection switch error (only
with S-STD93)
Turn the toggle switch to the left at the
bottom of the module.
5
Calibration not protected. (only with S-
STD93)
Protect calibration by selecting
Protection ON in the NIBP calibration
menu.
6
ADC error
ADC circuit failure. Change NIBP board.
7
Watchdog time too short
Change NIBP board.
8
Watchdog time too long
Change NIBP board.
9
Watchdog activated
Change NIBP board.
10
EEPROM checksum error; memory failure Change NIBP board.
11
Auto zero range exceeded
Calibrate NIBP.
12
Communication break; temporal break
down of communication from monitor
detected
Automatic recovery
13
-
-
14
Too early Auto Start (needs 25 seconds
without pressure)