UM-0085-B09
DT80 Range User Manual
Page 398
RG
Error Number – Description
•
Cause/Action
Error Category
Syntax
Operation
Memory
Reading
Hardware
E37 – No current job
•
A command was entered that operates on the current
job, but there is no job currently loaded.
E39 – Channel list fixed /F
•
Changes to program are not allowed
E42 – USB device not ready
•
No USB memory device inserted
•
DT80
has not yet read the required system information
from the device (wait a few seconds)
•
USB device is faulty or not a memory device
E49 – Job has logged data/alarms
•
A job that has logged data cannot be deleted. Delete
the data first.
E50 – Job locked
•
A job that has been locked cannot be modified
E51 – ALARM/IF command error
•
Setpoint character <, >, <> or >< missing
•
AND, OR, XOR incorrectly entered
•
Setpoint not specified or too large
•
Delay incorrectly specified
E52 – Text memory full
•
Memory for storage of alarm and expression text is full
E54 – Expression error
•
Syntax error
•
Expression too complex
E65 – ALARM undefined
•
Alarm n does not exist
E74 – Serial sensor CTS detect timeout
•
Serial sensor: CTS did not go to the required state
within timeout period
E75 – Serial sensor transmit timeout
•
Serial sensor was not able to transmit (e.g. due to flow
control state) within the timeout period
E80 – Serial device not responding
•
no response received from SDI-12 sensor
•
check cabling and sensor address
E81 – Serial device invalid response
•
garbled response received from SDI-12 sensor
•
check for address conflict
•
check for electrical noise
E82 – Serial device data not available
•
SDI-12 sensor doesn't support the requested
measurement
•
sensor doesn't support continuous mode
E89 – Serial sensor receive timeout
•
Serial sensor: expected characters were not received
within timeout period
E90 – Serial sensor scan error
•
Serial sensor: could not convert the received text as
specified in the control string
E101 – Undefined reference:
name
•
No channel was found with the specified name
E102 – Storefile record size too large
•
There are too many logged channels in the schedule
E103 – Invalid storefile:
filename
•
The indicated storefile appears to be corrupted
E104 – Drive format failed
•
Unable to format the specified drive. It may be write-
protected or damaged.