UM-0085-B09
DT80 Range User Manual
Page 124
RG
Part I – dEX
What is
dEX
?
The
DT80
provides an embedded
web interface (
dEX
)
that provides simple, intuitive access to the logger’s operations
using a standard web browser. You can view current sensor readings, job status and access data and other files stored
in the file system.
The standard web interface is built into the logger’s firmware. It does not need to be installed on the logger or on your
PC. Simply use your existing browser to browse to the logger’s IP address and the home page will be displayed.
Important:
dEX
version 1.84
covered in this manual
dEX
vs. Classic Web Interface
The
DT80
actually features two web interfaces:
•
dEX
, also known as the
enhanced web interface
, is available on all Series 2 or later loggers. This uses
Adobe
Flash
technology to provide a full-featured and easy to use interface.
•
The
classic web interface
is available on all loggers. This HTML-based interface has fewer features than the
enhanced interface, but it is customisable and is also more suitable for slow connections and small-screen web
clients.
The following table summarises the differences between the two web interfaces:
Feature
dEX
Classic web interface
Display basic logger information (serial number, firmware version etc.)
yes
yes
Display schedule status (log/run status, number of logged records, etc.)
yes
yes
Display current channel values in tabular form
yes
yes
Display event logs
yes
yes
Display user manual, release notes, etc.
yes
yes
Graphically configure the logger, i.e. define schedules, channels and global settings
without using the logger command language
yes
-
Display selected channels in a configurable "mimic" displays, including trend charts
yes
-
Unload data and save to CSV or DBD file
yes
-
Text command window (similar to
DeTransfer
)
yes
-
Display current job's program text
yes
-
Display file system free/used space breakdown
yes
-
Display TEST and SERVICEDATA command response
yes
-
User customisable branding (images, web links, help pages)
yes
-
Menu options can be selectively disabled or restricted
yes
-
Suitable for low bandwidth connection e.g. dial-up modem
-
yes
Suitable for small screen client e.g. PDA
-
yes
User customisable HTML pages
-
yes
Connecting to the Web Interface
In order to access the
DT80
web interface, a
TCP/IP
connection between the
DT80
and the PC is required. You can
either:
•
connect the
DT80
's Ethernet port to a local area network, or directly to the PC using an Ethernet cross-over
Ethernet Communications (P231)
)
•
use
DtUsb
to provide a TCP/IP connection via a USB cable (see
)
•
start a communication session using the integrated modem on DT8xM models (see
)
Once a TCP/IP connection has been established, all you need to do is type the
DT80
's IP address into the address field
of your browser. The web interface home page should then be displayed.
dEX
Home Page
For Series 1 loggers, when you enter the
DT80
's IP address into your browser you will be immediately directed to the
home page for the classic web interface (Series 1 loggers do not support
dEX
). See
for
more details.