UM-0085-B09
DT80 Range User Manual
Page 395
RG
Firmware Upgrade – USB Flash Device
The simplest way to upgrade the DT80 firmware is by using a USB memory device. This is also convenient if you have
several loggers to upgrade.
1.
Obtain the appropriate firmware upgrade zip file from
or your dataTaker representative.
This file should have a
–usbdisk
suffix.
2.
Extract all files from the zip file into the root directory of a USB memory device.
3.
If a USB or RS232 cable is connected to the logger, disconnect it.
4.
Ensure that the DT80 is powered up and awake, then Insert the USB memory device into the logger. The
oninsert.dxc
file on the USB device will then run. This will:
a. copy the firmware upgrade files to the DT80 internal memory (during this time the green
Disk
LED will
flash), then
b. start the firmware upgrade, which will proceed in a similar way to the USB/ RS232 method described
below.
5.
The logger will automatically restart once the upgrade is complete.
6.
Check the firmware version displayed on the LCD, or sends the TEST0 command.
Note:
If in step 4 the green
Disk
LED does not flash and instead the red
Attn
LED comes on, then there may be insufficient free
disk space on the logger. Check the available free space, as described above.
Firmware Upgrade – Host USB or RS232 Port
Here’s the procedure for upgrading the firmware of a DT80 by using
DeTransfer
(Version 3.18 or later) on a computer
directly connected to the DT80’s USB or Host RS 232 port:
1.
Obtain the appropriate firmware upgrade zip file from
or your dataTaker representative.
This file should not have a
–usbdisk
suffix. Extract all files from the zip file.
2.
Connect the host computer to the DT80 using USB. RS232 may also be used (57600 baud, 8 data bits, no
parity, 1 stop bit, software flow control)
Start
DeTransfer
.
3.
If you are using USB then the DtUsb driver software will need to be disabled in order to make the USB virtual
COM port available to
DeTransfer
. Open the DtUsb GUI by double clicking on the purple system tray icon, or
selecting DtUsb in the Windows start menu. Then click
Quit
to shut down DtUsb.
4.
In
DeTransfer
:
a. Open a connection to the DT80
b. Choose
Upgrade Firmware (DT80/800)
from the File menu.
c. Select the correct firmware file, then press OK
5.
The upgrade process will now proceed automatically. The DT80’s LEDs will flash and the LCD display should
indicate that the upgrade is progressing. You will notice that the upgrade proceeds through three phases:
a. A special "loader" program is downloaded to the logger (
Attn
LED flashes and display shows: DT8x
Bootstrap)
b. The main firmware package is downloaded to the logger (
Sample
LED flashes and display shows:
DT8x Loader). This package normally consists of a number of files (which are installed onto the
internal B: drive), followed by the firmware proper (which is installed into a special area of flash
memory).
Note:
If one or more of the files in the firmware package cannot be installed onto the DT80's internal drive (e.g. due
to insufficient disk space) then an error message (e.g.
fwrite: Cannot write data
) will be returned to the
host computer and the DT80 will restart. The firmware version will not have changed. Press the Details button on
the firmware upgrade window to see status and error messages returned by the logger.
c. A hard reset is performed (display shows the normal "sign-on" screen)
Important:
During the upgrade, do not remove any cables, or reset or power-down the DT80.
6.
Once the upgrade is complete, check that the version number displayed on the sign-on screen is correct. For
example if the file
DT80-9208771.dxf
was loaded then the display should indicate
DT80 V9.20
.
7.
Connect to the DT80 and re-load the required job if required.