Dakewe (Shenzhen) Medical Equipment Co., Ltd
77
Click the “Paraffin Replacement” or “Paraffin Filling” in the system interface
(Figure 71-1)
. The system will recalculate the paraffin melting time
(Figure
71-2)
. If adding liquid paraffin, press “Cancel” and the system will not
calculate the paraffin melting time.
Figure 71
It is recommended to refill the paraffin cylinder one by one.
If the paraffin cylinder is refilled with paraffin with different melting points, the
paraffin cylinder needs to be emptied and cleaned. After refilling, reset the melting
point of the paraffin cylinder.
If adding the paraffin particles, make sure that paraffin melts completely before the
program begins.