4.3.5 Exhaust Pipe (optional)
The instrument is designed to be connected to an external exhaust emission device. Put
the attached exhaust pipe (Figure 14-1) on an exhaust vent. Use a pipe clamp (Figure
14-2) to fix the exhaust pipe to the exhaust vent. The outlet of the other end of the
exhaust pipe is oriented outdoors or placed in a ventilation device.
Even if the instrument is connected to an exhaust emission device, the active carbon
filter equipped for HP300 Plus must still be installed.
4.3.6 Marble Workbench
An integrated marble workbench is mounted on the top of the HP300 Plus.
The workbench has characteristics of small adhesion and easy cleaning. It can
be wiped with a clean and lint-free cloth and, if necessary, any paraffin
dropped on it can be cleaned with xylene reagent.
The workbench area on the right side of the dehydration cylinder, can be used
to place the prepared samples and the tissue sample baskets (Figure 15-1)
removed from the dehydration cylinder.
4.3.7 Tissue Sample Basket
Figure 13
Figure 15