76
Dakewe (Shenzhen) Medical Equipment Co., Ltd
Pull the paraffin cylinder out of the paraffin case.
Pull out the paraffin cylinder slowly and carefully. Do not yank. The internal paraffin is
very hot and can cause burns.
The surface temperature of the paraffin cylinder is also very high. Therefore, when
replacing or adding paraffin, be sure to wear protective clothing, goggles, gas masks,
rubber gloves and all other necessary personal protective equipment.
Dispose of the used paraffin into a large container and refill the paraffin
cylinder. The marks of the highest and lowest filling liquid levels are located
on the inner wall of the paraffin cylinder.
Push the full paraffin cylinder back into the paraffin case. When inserting the
paraffin cylinder, make sure that the paraffin cylinder (male) is aligned with
the corresponding paraffin case joint (female). During insertion, a clicking
sound indicates that you have correctly installed the paraffin cylinder in place
(Figure 70)
. Close the paraffin case door afterwards.
Figure 70