32
Dakewe (Shenzhen) Medical Equipment Co., Ltd
the paraffin case and is used to collect spilled paraffin.
The waste paraffin collection tray needs to be inspected and cleaned regularly.
The waste paraffin collection tray can be removed from the mounting groove
by the handle to clean the waste paraffin inside.
4.3.10 Reagent Compartment and Reagent Bottle
HP300 Plus is equipped with 10 reagent bottles, 3 rinse bottles and 1
condensate bottle. These bottles are divided into two layers and placed in the
reagent compartment below the paraffin case (Figure 20-1). Each reagent
bottle is equipped with an RFID tag to identify and match the station and
reagent bottle (Figure 21-4).
The capacity of all bottles is 5L. There is a scale on the right front side of the
bottle. Additionally, the bottles are marked with the scale of the first 3L,
second 4.3L, third 4.8L liquid sensor located on the dehydration cylinder.
Each reagent bottle has two inlets and outlets. The lower ones are the liquid
inlet and outlet, and the upper ones are the gas inlet and outlet. The liquid inlet
and outlet are equipped with two seal rings (Figure 21-3).
Figure 20
Figure 18