Dakewe (Shenzhen) Medical Equipment Co., Ltd
11
2.2.5 Warning - Operation
The processor can only be operated by trained laboratory personnel according to the
intended use in this User Manual.
In case of emergency, turn off the main power supply and unplug the power cord.
Appropriate protective device to include lab coat, gloves and goggles must be worn when
operating the instrument.
Make sure that the instrument being operated is equipped with a strainer or exhaust pipe.
Even if the instrument is operated according to the specified purpose, dangerous solvent
flames can also be produced, which may harm the operator’s health and cause fire!
If need to interrupt the dehydration procedure for a long time, do not leave any tissue in
the processing cylinder to prevent the tissue from drying out.
2.2.6 Warning - Cleaning and maintenance
Before each maintenance or cleaning, the instrument shall be TURNED OFF and the
main power supply shall be disconnected. DO NOT use any solvents that contain acetone
or xylene to clean the instrument. Avoid liquid spillage on the internal components of the
instrument, including during operation and cleaning.
When using detergents, all safety requirements and laboratory management policies of
the product manufacturer shall be observed.
In any case, DO NOT flush the sites at high temperature. Otherwise, the sites may be
deformed!
Spilled solvents (reagents) must be wiped off immediately. Failing to do so could result in
corrosion of the processor!
In order to clean the lid, control panel and housing, use a mild household cleaner (such as
neutral soapy water).