background image

Instructions For Use

CA-9001 (Hose Assembly)

CA-9002 (Power Cord)

CA-1001 (Controller Starter Kit)

DA-101884 (Surgical Overlay)

DA-303584 (MedSurg Bed Overlay)

MC-303584 (MedSurg Bed Overlay-Micro)

Patent Pending

DOC# R01-0006-00004-AA  

Summary of Contents for DabirAIR CA-1001

Page 1: ...s For Use CA 9001 Hose Assembly CA 9002 Power Cord CA 1001 Controller Starter Kit DA 101884 Surgical Overlay DA 303584 MedSurg Bed Overlay MC 303584 MedSurg Bed Overlay Micro Patent Pending DOC R01 00...

Page 2: ...verlays 10 3 Setting up the System 3 1 Pneumatic Controller Pre Assembly 11 3 2 System Installation 12 14 4 Operating the Dabir System 15 5 Setting the Cycle Speed and Firmness Modes 5 1 Setting Cycle...

Page 3: ...handle Do not place Pneumatic Controller on top of loose articles which may block air intake ports System must be at room temperature prior to operation Allow Pneumatic Controller and Overlay to accli...

Page 4: ...or or moderate injury to the user or patient or damage to the product or other property This includes special care neces sary for the safe and e ective use of the device and the care necessary to avoi...

Page 5: ...to be used Keep Dry Indicates a medical device that needs to be protected from moisture Fragile Handle With Care Indicates a medical device that can be broken or damaged if not handled carefully Humi...

Page 6: ...that the equipment is suitable for direct current only ON power Indicates connection to the mains OFF power Indicates disconnection to the mains Universal Serial Bus Power Cord Box label FCC Declarat...

Page 7: ...s of Use The Dabir Surface Patient Support System is indicated for Assisting in reducing pressure ulcer development when combined with a comprehensive pressure ulcer prevention plan Reducing patient r...

Page 8: ...oped on the principle of supporting a person on small closely spaced areas of contact Comfort Nodes that dynamically alternate in height to relieve at risk tissue against pressure ulcer formation The...

Page 9: ...ow Medium High FIRMNESS DabirAIR Accessories 9 DabirAIR Hose Assembly Handle User Interface Low Medium High CYCLESPEED Low Medium High FIRMNESS Information Center Hose Assembly Connector Female Overla...

Page 10: ...climate Overlays 10 Overlay Assembly Connector Male Overlay Assembly Connector Male DOC R01 0006 00004 AA Micropressure Overlay vent located here CAUTION Keep Micropressure Overlay vent free and clear...

Page 11: ...rews into the Hose Assembly Connector Male and tighten with the provided Screwdriver CAUTION Do not over tighten the Attachment Screws CAUTION Do not use petroleum based lubricants on the Hose Assembl...

Page 12: ...ise speci ed in manufacturer s instructions Overlay and sheet must be at and wrinkle free CAUTION It is recommended that no additional layers be added between the patient and Dabir Overlay Additional...

Page 13: ...t away from the bed or support surface less than six 6 feet from a power outlet and in a place where the user has easy access to the control panel for operation CAUTION To ensure proper device groundi...

Page 14: ...r operation The device will not function if a positive connection is not made Disconnection Procedure Depress orange button and pull to release Overlay Assembly Connector Male CAUTION To avoid depress...

Page 15: ...rectly into an acceptable wall power outlet 2 Power Pneumatic Controller ON by depressing Main Toggle Switch see images below CAUTION Ensure Power Cord is clear of movable bed parts and placed to avoi...

Page 16: ...pendent of the START STOP feature by simply touching the new desired setting Information Center Low Medium High CYCLE SPEED Low Medium High FIRMNESS Information Center Low Medium High CYCLE SPEED Low...

Page 17: ...the key for the desired setting CAUTION To avoid bottoming out on the Overlay the MED or HIGH rmness settings are recommended for patients with greater BMI CAUTION To avoid bottoming out on the Overla...

Page 18: ...tended for use during CPR Audible Tones and Visual Indicators The Dabir Pneumatic Controller utilizes soft audible tones illuminated graphics and information displays to enhance the users operating ex...

Page 19: ...NOTE See the table on the following page for speci c error code messages and required action Error codes will automatically be cleared once the problem is resolved and operation is resumed by touchin...

Page 20: ...rced shutdown by the Pneumatic Controller Overlay must be replaced accordingly The Pneumatic Controller did not achieve pressure set point upon startup Power OFF using the Main Toggle Switch Check Ove...

Page 21: ...h a hospital grade disinfecting cleaner and water making sure to remove any organic debris Be sure to dry the Overlay thoroughly before using with patients Cleaner residue may e ect the Overlay slip r...

Page 22: ...d is plugged into the Pneumatic Controller and the appropriate wall outlet Blown fuse Contact manufacturer Internal malfunction Contact manufacturer Pneumatic Controller powers ON but does not operate...

Page 23: ...rage and Transport 40 C to 60 C 40 F to 140 F 10 95 RH Controller Speci cations Model Supply Voltage Supply Frequency Power Input Size Weight Case Material Fuse Ratings Type of protection against elec...

Page 24: ...ly Replacement Power Cord Model Application Width mm in Length mm in Weight kg lb Microclimate Air Flow Usable Life DA 101884 Surgical Table 2134 84 1 54kg 3 4 N 120 Hrs DA 303584 MedSurg Bed 2134 84...

Page 25: ...es Surge 1 kV line s to line s IEC 61000 4 5 2 kV line s to earth 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 11 Speci cation Sheets Continued Voltage dips short interruptions and...

Page 26: ...tters and electromagnetic site survey should be considered If the measured eld strength in the location in which the Model CA 1001 is used exceeds the applicable RF compliance level above the Model CA...

Page 27: ...and Mobile RF Communications Equipment and Model CA 1001 The Model CA 1001 system is intended for use in the electromagnetic environment speci ed below The customer or the user of the Model CA 1001 sy...

Page 28: ...r sent metteur radio peut fonctionner avec une antenne d un type et d un gain maximal ou inf rieur approuv pour l metteur par Industrie Canada Dans le but de r duire les risques de brouillage radio le...

Page 29: ...onger or for any additional period of time that may be required in the end user s jurisdiction This warranty does not cover problems or damage resulting from a accident abuse misapplication or any una...

Page 30: ...RPOSE OR AS TO ITS PERFORMANCE Any statement description or speci cation in Company s literature is for the sole purpose of identi cation of products sold by the Company and imparts no guarantee warra...

Page 31: ...ve Harwood Heights IL 60706 USA Corporate Headquarters Dabir Surfaces Inc 24585 Evergreen Road South eld MI 48075 USA Tel 1 248 796 0802 Fax 1 248 354 6667 Sales sales dabir surfaces com Technical and...