8 Cleaning Procedures
The cleaning procedures listed below are recommended by the manufacturer. Proper cleaning procedures
should be adjusted according to specific healthcare facility policy. Aggressive cleaning measures could
potentially damage system components.
CAUTION:
Turn Dabir System “OFF” during patient transfer and before patient positioning.
CAUTION:
If the Overlay is damaged in any way during cleaning, it must be replaced.
CAUTION:
Do not autoclave for cleaning.
Cleaning the DabirAIR Controller and Hose Assembly
Power “OFF” the Pneumatic Controller and disconnect the Power Cord from the wall outlet prior to clean
-
ing. Clean by wiping with a moistened cloth using mild soap and water. Clean all exposed areas. Remove
debris and allow the Pneumatic Controller to dry thoroughly.
CAUTION:
Do not saturate cloth or apply cleaning fluids/liquids directly to the Pneumatic
Controller
or Hose Assembly to prevent fluid ingress of this electronic device. Inspect air intakes
and exhaust ports located on the bottom side of the Pneumatic Controller and Handle End-plate
respectively to ensure they are not obstructed.
Cleaning the DabirAIR Micropressure and Microclimate Overlays
Disinfect by wiping down all areas with a hospital grade disinfecting cleaner and water, making sure to
remove any organic debris. Be sure to dry the Overlay thoroughly before using with patients. Cleaner
residue may effect the Overlay slip resistance. If this occurs, wipe the Overlay with a water moistened cloth
to remove the residue. Dabir Overlays are compatible with the following disinfecting cleaners or equiva
-
lent: Dispatch™ Hospital Cleaner – Clorox, Virex 256™ - Diversey, Isopropyl Alcohol.
Ensure there are no sharp objects between the Overlay and patient support surface, mattress or pad.
9 Routine Maintenance
The Dabir Surfaces Patient Support System
It is recommended that the
Pneumatic Controller
be returned to the manufacturer for annual
maintenance and performance check after every 12 months of use. Routinely check all electrical
connections, the Power Cord and Hose Assembly for signs of wear or damage and replace if appli
-
cable with manufactured certified replacement parts. For any non-serviceable damage noted,
please contact the manufacturer or certified representative.
Serial Number Labels
The Pneumatic Controller serial number is located on the underside label.
The Hose Assembly serial number is located on the Hose Assembly label.
The Overlay serial number is located on the Overlay label.
All serial numbers are in the following format: LSYYDDDXXXXT
Please provide component serial numbers when requesting service.
DOC# R01-0006-00004-AA