7 Information Center: Error Codes and Alarms - Continued
Information Center
Buzzer
Meaning
Indicator
The Pneumatic Controller is not properly communicating
with the Overlay. Check the Hose Assembly to Overlay
connection. Light will clear when proper connection is
made and the unit will return to normal operation. If light
does not clear, change the Overlay. If light still does not
clear, contact the manufacturer or certified representative.
The Pneumatic Controller is not properly communicating with
the Hose Assembly. Check the Hose Assembly connection
to the
Pneumatic Controller. Power the unit “OFF” and wait
5 seconds using the main toggle switch. Power the unit
“ON” again. If the light does not clear, please contact the
manufacturer or certified representative.
The Overlay has exceeded its usable life or its allowable
shelf life and must be replaced. System will continue to
run for 12hrs or until the unit is stopped or powered “OFF”.
The Overlay is within 24 hours of its usable life and within
36 hours of a forced shutdown by the Pneumatic Controller.
Overlay must be replaced accordingly.
The Pneumatic Controller
did not achieve pressure set
point upon startup. Power “OFF”using the Main Toggle
Switch. Check Overlay and Hose Assembly for possible
leaks. Power the
Pneumatic Controller back “ON” again.
If problem persists, please contact the manufacturer or
certified representative.
20
Check
Connections
Continuous
Check
Connections
Flashing
2sec ON – 2sec OFF
Replace Overlay
Continuous
Replace Overlay
Flashing
1sec ON – 3sec OFF
Service Unit
Flashing
2sec OFF – 2sec ON
Service Unit
Continuous
Frequency: 2kHz
2sec ON
after touching
START/STOP button, then
OFF continuous
Frequency: 2kHz
2sec ON
after touching
START/STOP button, then
OFF continuous
Frequency: 2kHz
1sec ON – 30sec OFF
Frequency: 2kHz
2sec ON after 24hr mark,
then OFF continuous
Frequency: 2kHz
3x - 0.5sec ON, 0.5sec
OFF, then
OFF continuous
Frequency: 2kHz
3x - 0.5sec ON, 0.5sec
OFF, then
OFF continuous
The Pneumatic Controller
has encountered a physical error.
Power “OFF” using the Main Toggle Switch. Wait 5 seconds
and power the
Pneumatic Controller back “ON” again. If the
problem persists, please contact the manufacturer or
certified representative.