25
Model CA-1001 Manufacturer’s Declaration – Electromagnetic Immunity
The
Model CA-1001
system is intended for use in the electromagnetic environment specified below. The
customer or the user of the
Model CA-1001
system should assure that it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level
Electromagnetic Environment/Guidance
Electrostatic ± 6 kV contact
Floors should be wood, concrete, or
discharge
(ESD)
Josh
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should
be
at
least
30%.
IEC 61000-4-2 ± 8 kV air
Electrical fast ± 2 kV for power
transient/burst
supply
lines
IEC 61000-4-4 ± 1 kV for
input/output lines
Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5 ± 2 kV line(s) to earth
< 5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
11 Specification Sheets - Continued
Voltage dips,
short
interruptions
and voltage
variations on
power
supply input
lines
IEC 61000-4-11
Mains power quality should be that of a
typical commercial or hospital environ-
ment. If the user of the
Model CA-1001
requires continued operation during
power mains interruptions, it is recom
-
mended that the
Model CA-1001
be
powered from an uninterruptible power
supply or a battery.
70% UT
(30% dip in UT)
for 25 cycles
< 5% UT
(>95% dip in UT)
for 5 s
NOTE:
UT is the AC mains voltage prior to application of the test level.
DOC# R01-0006-00004-AA
Mains power quality should be
that of a typical commercial
or hospital environment
Mains power quality should be
that of a typical commercial
or hospital environment
Summary of Contents for DabirAIR CA-1001