3.2 System Installation
Overlay Installation
1.
Place Overlay flat onto patient support surface, mattress or pad with “THIS SIDE UP”
facing up. The hose routing position is preferred to be at the foot end.
CAUTION:
Ensure there are no sharp objects between the Overlay and patient support surface,
mattress or pad. Avoid placement near sharp or hot objects to avoid puncture.
CAUTION:
If Overlay is damaged or punctured in installation and/or usage, it must be replaced.
CAUTION:
Never place Overlay directly on bed frame. To assure proper function, the Overlay
must be placed on top of a support mattress or pad.
3.
Place facility supplied, standard cover sheet over the top of the Overlay and tuck edges
underneath the mattress or pad unless otherwise specified in manufacturer’s instructions.
Overlay and sheet must be flat and wrinkle free.
CAUTION:
It is recommended that no additional layers be added between the patient and Dabir
Overlay. Additional layers may negativly impact performance.
CAUTION:
At the care-giver's discretion, Dabir Surgical Overlays (DA-101884) may be used without
bedsheets for a single surgical procedure.
12
NOTE:
Latex is not used in the manufacture of this product.
Cover Sheet
Dabir Overlay
Mattress
Installation Cross Section
2.
Power “ON” the Pneumatic Controller and touch the cycle “START/STOP” key. Inspect the
Overlay to insure that all nodes are properly inflating and deflating. If the Overlay nodes
are not inflating properly, inspect the Overlay and Hose Assembly routing to insure that
there are no kinks restricting air flow.
Cover sheet to be placed
above the Dabir Overlay.
(Not shown)
DOC# R01-0006-00004-AA
MedSurg Bed Installation
Dabir Overlay
(Gray side down)
Summary of Contents for DabirAIR CA-1001