Angel
®
cPRP System
Operator’s Manual
7-2
Technical Data
Environmental Limitations
Characteristic
Performance
Condition
Mains Voltages
100 - 240 V~
Mains Current
5A
Maximum
Line Frequency
50 Hz-60 Hz
Ambient Operating
Temperature
10 to 30 C (50 to 86 F)
Ambient Operating
Humidity
10 to 75%
Non-condensing
Storage Temperature
-20 to 37.7C (-4 to 100 F)
Storage Humidity
10 to 90%
Non-condensing
Restrictions
Not to be used in an
explosive atmosphere
Emission test
Compliance
Electromagnetic
environment -
guidance
RF emissions CISPR 11 used for domestic Group 1 The Angel
®
Concentrated
Platelet Rich Plasma
(cPRP) System uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low
and are not likely to cause
any interference in
nearby equipment.
RF emissions CISPR 11 A
The Angel
®
Concentrated Platelet
Rich Plasma (cPRP)
System is suitable for
use in all establishments
other than domestic
and those directly
connected to the public
low-voltage power supply
network that supplies
buildings purposes.
Harmonic emissions
IEC 61000-3-2
A
Voltage fluctuations /
flicker emissions IEC
61000-3-3
Complies
Guidance and manufacturer’s declaration – electromagnetic emissions
The Angel
®
Concentrated Platelet Rich Plasma (cPRP) System is intended for
use in the electromagnetic environment specified below. The customer or the
user of the Angel
®
Concentrated Platelet Rich Plasma (cPRP) System should
assure that it is used in such an environment.