Angel
®
cPRP System
Operator’s Manual
2-3
Installing the Disposables
processing cycle. The PPP Compartment is used to collect platelet poor plasma;
the PPP is the first blood component collected after separation has been completed.
Syringe-activated valves are used to access the PPP and whole blood
compartments of the three-compartment reservoir bag.
Other accessory items in the Angel
®
cPRP Processing Set are:
20 ml Luer Lock Syringe: The 20 ml luer lock syringe is used for the
collection of platelet rich plasma. However, the syringe-activated PRP valve will
accommodate most luer fitting syringes.
60 ml Specimen Cups (2 ea.): Two 60 ml specimen cups for use in a sterile field.
Male/Female Luer Plugs: The male / female luer plugs are used during and at
the end of procedure to seal open luer lock connections.
Whole Blood Bag Spike Adapter: The whole blood spike adapter is used to
transfer the blood or mixture of blood and bone marrow from a whole blood bag
to the whole blood compartment of the three-compartment reservoir bag.
Labels: Appropriate labels to label collected whole blood/bone marrow and
separated components.
Contents of this set have been sterilized by ethylene oxide gas and have
nonpyrogenic fluid pathways.
Please see “Warnings” on page ii in the chapter “Before You Get Started”; see
“Precautions” on page iv in the same chapter.
Warnings and Precautions
1. Turn on the Angel
®
cPRP System be pressing the power switch on the back
of the machine. The message “Self test in progress. Please stand by.” will
be displayed on the Angel
®
cPRP System’s touch screen display, and the
machine will orientate the valve assembly driver to the loading position.
Turning on the Angel
®
cPRP System
Setup and Blood/Bone Marrow Preparation